Overview:

In a large-scale retrospective validation involving over 1,700 consecutive patients presenting to German Chest Pain Units, the PMcardio STEMI AI ECG Model demonstrated high diagnostic accuracy (95.3%) and specificity (96.7%) in identifying high-risk patients typically missed by standard ECG interpretation. This highlights its potential to significantly improve emergency care through earlier detection and intervention in a real-world, broad clinical setting.

Presented at: German Society for Cardiology (DGK) 90th Annual Congress
Presented Date: April 10, 2024

Background

Occlusion myocardial infarction (OMI) might be missed by conventional ST-elevation myocardial infarction (STEMI) criteria on 12-lead surface electrocardiogram (ECG). Twenty-five percent of non-STEMI patients have OMI in worldwide registries. These patients are at risk for delayed invasive management resulting in poor short- and long-term prognosis. 

Artificial intelligence (AI) model studies have shown robust performance in detecting angiographically determined OMI using individual 12-lead surface ECGs. However, these studies were conducted in high-risk patient populations with a high prevalence of OMI and more data is needed to evaluate the utility of AI in a broader chest pain (CP) population.

We investigated the performance of an AI ECG model detecting angiographic OMI in a German metropolitan chest pain unit (CPU) cohort of all-comer patients without typical STEMI criteria from a tertiary care center.

Methods

The AI model was developed and validated in international cohorts including >12,000 CP patients with an OMI prevalence of around 20%. The present study included all consecutive patients presenting to CPU who were either ruled out by serial high-sensitive cardiac troponin T (hs-cTnT) testing or referred for angiography after CPU assessment.

STEMI patients diagnosed in the prehospital phase were excluded from the analysis (Figure 1). In Germany, peak troponin is not routinely measured. Consequently, an angiographic OMI outcome was modeled including identification of culprit lesion, coronary artery flow and hs-cTnT.

The AI model performance was assessed according to sensitivity, specificity, accuracy and predictive values in the overall cohort and in the subgroup of ruled-out population following 0/1h hs-cTnT algorithm. Receiver operating characteristic (ROC) analyses were performed.

Results

In total, 1776 ECGs from 1770 CPU patients were considered. The mean age of patients was 53.2 (±18.9) years, and 62.1% were male. Eighty percent of patients were ruled-out by serial hs-cTnT measurement, and the remaining 354 patients were referred for invasive management.

OMI prevalence was 3.16%. In the overall analysis, the AI ECG model achieved an area under the curve of 0.894 in identifying OMI (Figure 2 and table 1). Sensitivity was 0.518 (95% CI: 0.386-0.649), and specificity reached 0.967 (95% CI: 0.959-0.976). Correspondingly, negative predictive value (NPV) was 98.4% (95% CI: 97.8-99.0), and positive predictive value (PPV) was 34.1% (24.2-44.4). Diagnostic accuracy was 0.953 (95% CI: 0.943-0.963).

Subgroup analysis of patients ruled-out following 0/1h hs-cTnT algorithm identified misclassification as false-positive in 1% of ECGs by AI model. Specificity and accuracy both were 0.989 (95% CI: 0.983-0.994, see table 2). Negative predictive value was 100% (95% CI: 100-100).

Conclusion

This external validation study of a pre-existing in AI ECG model demonstrates remarkable diagnostic accuracy in a German CPU cohort without a pre-clinical diagnosis of STEMI. The AI model was reliable with high specificity and resulted in overcalls only in few cases.

Sensitivity analysis of the AI ECG model was restricted by the low prevalence of OMI and by the modeled OMI definition given the German CPU specifics. The present study proved technical feasibility, but future studies are required to further evaluate its applicability in German CPUs.


Authors: Sascha Macherey-Meyer (Köln), P. Meyers (Charlotte, NC), S. Smith (Minneapolis), F. Hundehege (Köln), L. Krücken (Köln), S. Heyne (Köln), M. Meertens (Mainz), S. Braumann (Köln), V. Mauri (Köln), S. Baldus (Köln), S. Lee (Köln), C. Adler (Köln)

Author-Logo_PM
Powerful Medical leads one of the most important shifts in modern medicine by augmenting human-made clinical decisions with artificial intelligence. Our primary focus is on cardiovascular diseases, the world’s leading cause of death.

About PMcardio

PMcardio is a CE-certified AI that reads ECGs and offers a complex assessment of 49 cardiac conditions. Clinically validated in 15+ studies and trusted by over 100,000 clinicians, it delivers rapid, expert‑level interpretations, empowering emergency physicians, GPs, nurses, paramedics, and cardiologists to act with confidence at the point of care. Available for Individuals and Organizations.

About Powerful Medical

Established in 2017, Powerful Medical has embarked on a mission to revolutionize the diagnosis and treatment of cardiovascular diseases. We are a medical company backed by 28 world-class cardiologists and led by our expert Scientific Board with decades of experience in daily patient care, clinical research, and medical devices. The results of our research are implemented, developed, certified, and brought to market by our 50+ strong interdisciplinary team of physicians, data scientists, AI experts, software engineers, regulatory specialists, and commercial teams.

Share this article

Relevant Publications

Can an AI ECG algorithm improve diagnostic accuracy for acute coronary occlusion in the difficult subset of canceled catheterization lab activations?

Discordance in ECG interpretation between Emergency Medicine and Cardiology teams is common, and within canceled STEMI activations, a true acute coronary occlusion myocardial infarction (OMI) can go unrecognized. This retrospective study examined whether an AI ECG algorithm (Queen of Hearts™) could improve OMI detection in this difficult subset. Across three referral centers, the investigators analyzed 185 activations canceled for not meeting STEMI criteria, of which 17 met the definition of a missed OMI. The AI algorithm identified 16 of 17 cases, far exceeding STEMI criteria in sensitivity (94.1% vs. 47.1%), supporting its use as an adjunct to clinical judgment in ambiguous cases.

Artificial Intelligence-Assisted, ECG-Based Triage of Patients With Chest Pain to Immediate Invasive Treatment

Rapid identification of acute coronary occlusion (ACO) is critical in chest pain patients, yet conventional STEMI criteria miss occlusion in many NSTEMI cases, delaying life-saving invasive treatment. This retrospective study tested whether a deep learning ECG AI model could improve ACO detection in an unselected cohort of more than 4,000 consecutive patients from a German chest pain unit. Each 12-lead ECG was assessed using both standard STEMI criteria and the AI model, with ACO independently adjudicated by a blinded physician. The AI model clearly outperformed STEMI criteria, identifying 73 of 104 ACO cases versus 30 (area under the curve 0.958 vs. 0.589), with fewer false positives. The findings suggest AI-assisted ECG interpretation can detect subtle ischemic changes beyond established criteria and support faster triage of NSTE-ACO patients to immediate invasive care.

Join over 100,000 healthcare professionals who are already taking advantage of AI